تعداد نشریات | 50 |
تعداد شمارهها | 2,232 |
تعداد مقالات | 20,475 |
تعداد مشاهده مقاله | 25,235,095 |
تعداد دریافت فایل اصل مقاله | 22,869,506 |
Chromium (III) ion selective electrode based on di(benzylamino)glyoxime | ||
Journal of Physical & Theoretical Chemistry | ||
مقاله 6، دوره 3، شماره 2، آبان 2006، صفحه 43-48 اصل مقاله (240.75 K) | ||
نوع مقاله: Research Paper | ||
نویسندگان | ||
M. Kia* ؛ H. Aghaie* ؛ M. Arvand* ؛ K. Zare* ؛ M. Aghaie* | ||
Not-mentioned | ||
چکیده | ||
A new poly(vinylchloride) membrane sensor for Cr3+ ions based on di (benzylamino)glyoxime as an ionophore was prepared. The electrode has a linear dynamic range 1×10-6-1×10-1 mol l-1,with a Nernstian slope of 20.3 ± 0.5 mV per decade and a detection limit of 2×10-7. It has a fast response time of <15 s and can be used for at least 3 months without any considerable divergence in potential. The proposed electrode revealed good sensitivity for Cr(III) over a wide variety of metal ions and could be used in a pH range of 2-5.5. The proposed sensor was used for the determination of Cr3+ ions in real samples such as tea leaves, coffee and cacao powder by direct potentiometry. | ||
کلیدواژهها | ||
Ion-selective electrode؛ Chromium(III)؛ Di(benzylamino)glyoxime؛ Poly(vinyl chloride) membrane | ||
آمار تعداد مشاهده مقاله: 721 تعداد دریافت فایل اصل مقاله: 324 |